Nous utilisons des cookies pour offrir aux visiteurs de notre site web la meilleure expérience possible. Si vous continuez sans changer les paramètres de votre navigateur, nous supposerons que vous êtes heureux de recevoir tous les cookies sur le site web de BioTek. Pour en savoir plus concernant notre utilisations de nos cookies ou comment faire pour bloquer, veuillez consulter notre politique, visite notre cookie policy.

Fluorescence Intensity

Fluorescence intensity detection has a much broader range of applications than absorbance detection. For fluorescence intensity measurements, an optical system (the excitation system) illuminates the sample using a specific wavelength (selected by an optical filter, or a monochromator), thereby exciting the sample. The excitation causes the sample to emit light (i.e. fluoresce) at a different wavelength. The emitted light is collected by a second optical system (emission system) and the signal is measured by a light detector such as a photomultiplier tube (PMT).

Samples are chemically bound with fluorescent tag

Samples are chemically bound with fluorescent tag

Light at a specific wavelength excites the sample

Light at a specific wavelength excites the sample 

Excited samples emit light at a different wavelength

Excited samples emit light at a different wavelength

The advantages of fluorescence detection over absorbance detection are increased sensitivity and a broader application range, given the wide selection of fluorescent labels available today. For example, a technique known as calcium imaging measures fluorescence intensity of calcium-sensitive dyes to assess intracellular calcium levels.